Automated lab safety system

Fully automated closed position control with optional leak testing for gas withdrawal points in school and laboratory facilities

Safe gas supply for laboratories in schools and research facilities – standard-compliant, reliable, automated.

Fully automated gas shut-off

Before each use, the system checks whether the gas withdrawal points in the classroom or laboratory are closed and only enables the gas supply once this has been confirmed.

Reliable protection against gas accumulation and explosions

The safety valve consists of a dual solenoid valve with an integrated pressure sensor and prevents gas leaks. In the event of a fault – such as a gas leak or a drop in pressure – it automatically and reliably shuts off the gas supply.

Easy to operate

The system is operated directly on the LCU 300 Compact itself or via push-buttons on the connected remote control FB 300. All operating states are displayed clearly on the remote control and indicated by a multicolour LED display on the unit. An overview of the meaning of the LEDs is provided on the housing cover.

Standard-compliant & certified

Meets the requirements of DVGW worksheet G 621 and DIN 30666 – ideal for schools, laboratories and research facilities.


Safety for teachers, students and laboratory staff

Bild LCU300

Before each use, the lab control unit automatically checks whether all gas withdrawal points are closed and whether the required inlet pressure is available. The gas supply is only enabled once this check has been passed successfully.

The control unit can operate two different valves: a central shut-off device VCC without closed position control, and either a laboratory safety valve VCL with dual solenoid valve or a laboratory safety valve VAL with one solenoid valve, each with closed position control in accordance with DVGW worksheet G 621 (A).

Using the integrated DIP switch, the automated lab safety system can be configured flexibly and, if required, extended with optional leak testing. This enables recurring leak tests of flexible lines, as required by DIN 30666.


Can be used anywhere flexibly – from the classroom to the research lab

Unterrichtsraum Labor

Classrooms

Ensures a safe and standard-compliant gas supply in the classroom.

Bunsenbrenner mit Flamme

Teacher & student benches

Ensures safe use through the separate monitoring of the withdrawal points for teachers and students.

Forschungslabor

Research labs

Monitors the gas supply in demanding laboratory environments.

Who is the automated lab safety system designed for?

Schule

Schools, laboratories & technical training rooms

Learning, research and experimentation at all times under safe conditions.

Labormontage

Professional laboratory fitters & furniture makers

Individual configuration via DIP switches. Multicolour LED display for operating states and error messages.

Technische Zeichnung

Planners & engineering firms

Compliant with DVGW worksheet G 621 (A) and DIN 30666. Reliable for planning and tested.

Automated lab safety system - according to DVGW worksheet G 621 (A)

Automated lab safety system for gas withdrawal points in laboratories, classrooms and technical workrooms in accordance with DVGW worksheet G 621 (A) and DIN 30666, to ensure and monitor a safe gas supply in laboratory environments.

Automated lab safety system consisting of:

  • Lab control unit LCU
  • Central shut-off device VCC as a shut-off valve upstream of the teacher bench withdrawal point
  • Laboratory safety valve VCL with closed position control and optional leak testing upstream of the student bench withdrawal points


Components of the lab safety system type LCU


Technical data & downloads

TypeLCU 300
Operating voltage230 V ~ / 50–60 Hz
Power consumptionmax. 90 VA
Protection typeIP 54
Ambient temperature0 - 60 °C
HousingSurface-mounting, plastic, light grey (RAL 7035), clear plexiglass cover
DisplayRGB-LED for operating- and fault statuses
OptionalLeak testing according to DIN 30666
CertificationDIN EN 13611, EU/2016/426, EC type-tested
Valve versionsVCC = Central shut-off device
VCL = Laboratory safety valve with dual solenoid valve
VAL = Laboratory safety valve with one solenoid valve
Series3 = series 3
4 = series 4
Nominal diameter15 = DN 15
20 = DN 20
25 = DN 25
ConnectionR = female thread Rp
VH = female thread Rp with brass screw connection
Inlet pressure3 = 360 mbar
5 = 500 mbar

Contact for laboratory fitters, furniture makers & planners

Are you planning a laboratory or operating a gas laboratory installation? Contact us for individual advice and project planning.

Sales Representative

Mr. Karl-Heinz Rahner

Region: Germany (excluding Bavaria and Baden Württemberg), Austria, Switzerland and Luxembourg


Mobile: +49 151 74308530
E-mail: k.rahner@gok.de

Sales Representative

Mr. Andreas Lauck

Region: Bavaria and Baden-Württemberg


Mobile: +49 170 6412794
E-mail: a.lauck@gok.de

Business Consulting

Telephone: +49 9332 404-9292
E-mail: kundendienst@gok.de

Technical Support

Telephone: +49 9332 404-9393
E-mail: support@gok.de


Find a specialist distributor in your area

Would you like to order the EMS exhaust gas extraction monitoring system or the KCU kitchen exhaust safety system? Our specialist distributors will support you with their extensive expertise, advise you on selecting the right version and reliably take care of the entire ordering process.

You haven't found a suitable contact in your area yet, or would you prefer to coordinate directly with GOK? We will be happy to assist you, from product advice through to order processing, if desired in close coordination with a specialist distributor. Simply contact us by email at kundendienst@gok.de or by telephone on +49 9332 404-9292.